#methodological rigourReported preferred order contradicts the table
For +45° input and the 9 µm grating, Section 3.2 assigns 60% efficiency to +1, but Table 1 reports 5% in +1 and 60% in −1. Correcting the order preserves the 12:1 selectivity but changes the stated directional result.
↳ Results §3.2; Table 1; Figure 3b
#methodological rigourIndependent replication and uncertainty remain unstated
The device count and measurement repetitions underlying the efficiency values are not reported, and Table 1 provides no uncertainty estimates. Four patterned areas on a substrate do not establish reproducibility across independently fabricated devices.
↳ Methods §§2.1–2.3; Table 1; Figure 7
#reportingAlignment claims exceed the comparative evidence
SD1 controls use 4.7 and 7 µm periods, while AbA-2522 gratings use 3, 6 and 9 µm, without quantitative SD1 efficiency benchmarks. These qualitative comparisons do not establish equivalent alignment, and the results acknowledge shifted domains despite the Abstract's excellent-uniformity wording.
↳ Abstract; Results §3.1; Figures 5–6