#methodological rigourDetection uncertainty lacks independent repeat characterization
The concentration-series SNR plot gives no independent repeat counts or error bars, and its RMSE denominator is undefined. Pulse averaging does not establish variability across repeated fills, alignments or phantom preparations.
↳ Results, concentration series; Fig. 5e
#methodological rigourPrepared surfaces restrict the demonstrated applicability
Detection uses clear, glass-molded PDMS with a specular surface and opposite-side illumination. These conditions do not establish performance on rough, scattering biological samples.
↳ Methods, Silicone Sample; Fig. 1a; Conclusion and Outlook
#methodological rigourImage performance remains only partly quantified
The approximately 230 µm resolution is theoretical rather than measured, and channel-localization error is not quantified. Realignment and illumination differences also confound amplitude comparisons between depths.
↳ Results, Non-contact Photoacoustic Imaging; Fig. 3e