On-chip flow cytometer using integrated photonics for the detection of human leukocytes.

Jooken et al.Jooken, Zinoviev, Yurtsever et al.·2024·Scientific reports·10.1038/s41598-024-60708-0 ↗

Primary claim

The optical device is capable of detecting the side- and forward-scattered light of white blood cells, with sufficient resolution for the discrimination of monocytes and lymphocyte populations in a peripheral blood mononuclear cell (PBMC) sample.

Supported · complementary benchmark and cell-subset measurements in a single donor
In brief

Integrated photonics enables lymphocyte–monocyte discrimination against a commercial benchmark, but single-donor testing confines the result to a proof of concept.

ReliabilityCaveats4 of 4 checks completeRead the entry ↓
Fragile — the claim's weakest requirement is that it holds beyond the sample studied. Based on how the study was designed and conducted, not on statistics extracted from the paper. Experimental. Every point below is traceable to a quoted passage, but this forecast has not yet been validated against known replication outcomes.
Abstract

Differentiation between leukocyte subtypes like monocytes and lymphocytes is essential for cell therapy and research applications. To guarantee the cost-effective delivery of functional cells in cell therapies, billions of cells must be processed in a limited time. Yet, the sorting rates of commercial cell sorters are not high enough to reach the required yield. Process parallelization by using multiple instruments increases variability and production cost. A compact solution with higher throughput can be provided by multichannel flow cytometers combining fluidics and optics on-chip. In this work, we present a micro-flow cytometer with monolithically integrated photonics and fluidics and demonstrate that both the illumination of cells, as well as the collection of scattered light, can be realized using photonic integrated circuits. Our device is the first with sufficient resolution for the discrimination of lymphocytes and monocytes. Innovations in microfabrication have enabled complete integration of miniaturized photonic components and fluidics in a CMOS-compatible wafer stack. In combination with external optics, the device is ready for the collection of fluorescence using the on-chip excitation.

Key Claims

Primary finding · Supported

The optical device is capable of detecting the side- and forward-scattered light of white blood cells, with sufficient resolution for the discrimination of monocytes and lymphocyte populations in a peripheral blood mononuclear cell (PBMC) sample.

Fragile — the claim's weakest requirement is that it holds beyond the sample studied. Based on how the study was designed and conducted, not on statistics extracted from the paper. Experimental. Every point below is traceable to a quoted passage, but this forecast has not yet been validated against known replication outcomes.

complementary benchmark and cell-subset measurements in a single donor

Introduction, final device-capability paragraph
Quote verified
Secondary · Tentative

the composition of the cell population as found on-chip, 15.7% monocytes, and 77.4% lymphocytes, corresponds well to the composition given by the benchmark flow cytometer, 22.6% and 75.9%.

single-donor comparison with a 6.9-percentage-point monocyte difference and no quantified agreement uncertainty

Results, Detection of full PBMC populations; Fig. 6
Quote verified
Secondary · Supported

On-chip, 95.3% of the CD3 + cells within Q1 fall within the lymphocyte gate, compared to 94.5% on the benchmark flow cytometer.

single-donor back-gating restricted to detected CD3-positive cells

Results and Discussion, Fluorescence, paragraph 1
Quote verified
Secondary · Supported

the on-chip discrimination factor of FSC in ALL mode amounts to 1.4, which is on par with the discriminating factor of 1.4 obtained on the benchmark flow cytometer.

matching measured discrimination factors in isolated populations from one donor

Results, Detection of isolated PBMC populations
Quote verified

Derived from the full evaluation — not a separate score.

Strengths · Limitations

Strengths

#contribution

Integrated scatter collection distinguishes leukocyte populations

Monolithically integrated illumination and collection optics resolve lymphocyte and monocyte populations in the tested PBMC sample. This extends the bead-discrimination and counting capabilities described for earlier integrated devices.

↳ Introduction; Results, Figs. 5–6

#methodological rigour

Parallel benchmarking anchors the device measurements

Every sample was measured on a commercial CytoFLEX S instrument. Calibration beads, isolated leukocyte populations and CD3 fluorescence back-gating provide complementary checks on the on-chip signals.

↳ Methods, Benchmark cytometry measurements; Results, Figs. 3 and 5–7

#positioning

Physical limitations connect directly to performance

The analysis links limited collection efficiency and particle-height variation to sensitivity and population spread. Two interrogation positions test the proposed contribution of inertial focusing to bimodal bead side-scatter signals.

↳ Results, Fig. 4; Conclusions and perspectives

Limitations

#methodological rigour

Single donor limits biological performance validation

All PBMC experiments use aliquots from one donor, and three repeated measurements do not establish donor-to-donor performance. Chip-to-chip reproducibility is also unquantified.

↳ Methods, Cell preparation; Results, Detection of full PBMC populations

#reporting

Monocyte agreement wording exceeds the comparison

The on-chip monocyte fraction is 15.7%, compared with 22.6% on the benchmark: a 6.9-percentage-point difference, approximately 31% relative. Describing these compositions as corresponding well understates the unresolved difference.

↳ Results, Detection of full PBMC populations; Fig. 6

#methodological rigour

Fluorescence checks only part of classification

CD3-positive back-gating checks a subset of lymphocytes rather than all lymphocytes or monocyte identity. High fluorescence background and zero assignment to undetected fluorescence events further limit this validation.

↳ Results, Fluorescence; Fig. 7

Does this match your read? Add your assessment ↓

Nabu Commentary

Integrated illumination and scatter collection distinguish lymphocyte and monocyte populations, with parallel commercial measurements and complementary calibration experiments supporting the proof of concept. The physical characterization connects particle-position effects and collection geometry to observed performance. Single-donor testing and incomplete cell-identity validation prevent broader conclusions about classification accuracy or reproducibility. The full-PBMC monocyte fractions also differ more than the agreement wording suggests, although the numerical results are openly reported.

Evaluate another paper

Free · About 5 minutes · Research preview

Author Note / Field Feedback

Nabu’s assessment, alongside the field’s view.

Are you an author of this paper?

Dimension breakdown

Evidence Score

Out of 5
012345
3.7±0.2

Contribution
3.8

The device extends integrated photonic cytometry from bead discrimination or counting to leukocyte-subpopulation discrimination using integrated illumination and scatter collection. The demonstrated advance remains a single-donor proof of concept rather than a broadly validated classification platform.

Introduction; Results, Figs. 5–7; Methods, Cell preparation
Methodological Rigour
3.5

Parallel commercial measurements, calibration beads, isolated populations and CD3 back-gating provide complementary support for the detection result. Single-donor testing, unquantified chip-to-chip reproducibility and subset-specific fluorescence validation limit the performance conclusions.

Methods, Benchmark cytometry measurements and Cell preparation; Results, Figs. 3–7
Reporting
3.7

The experimental sequence and distinction between axial light-loss and true forward scatter are explicit. Agreement wording in the full-PBMC comparison is more favorable than the reported monocyte fractions warrant.

Results, Collection photonics and Detection of full PBMC populations

"corresponds well"

Positioning
3.8

The paper distinguishes earlier integrated approaches and traces optical collection and particle-position limitations to sensitivity and population spread. Quantitative positioning against other devices and discussion of biological generality are less developed.

Introduction, prior-device discussion; Conclusions and perspectives

Reliability

Caveats4 of 4 checks


Results checkClean

The reported numbers and statistics are consistent with the stated results.

Reporting and transparencyNoted

The supplied text lacks an ethics, waiver or consent declaration for the human biobank material and contains an incorrect Figure 7 panel reference. Commercial funding, competing interests and request-based data access are disclosed.

Conflict of interest: declared · Ethics approval: not stated · Data availability: declared

Reference IntegrityClean

29 references in manuscript 29 of 29 checkable references found in an index

References not verified automatically are resolved manually within 24 hours.

Post-Publication RecordClean

No retraction or expression of concern found in Crossref (includes Retraction Watch).

Last checked: 25 Nov 2026

Sources: Crossref (incl. Retraction Watch) ✓

Impact Profile

Impact areaHealth · Generic health relevance
Type of advanceMethodological
Route to useNamed
AdvanceAdvances
CitationsFWCI 1.0
Policy · patentNot yet traced
ReachedNot reachedNot assessable or unverified
ReadinessPiloted
Foundational
Controlled
Piloted
Validated
Implementable
FoundationalImplementable

The device is a proof of concept, demonstrated in one lab on PBMC samples from a single donor and benchmarked against a commercial cytometer, without independent confirmation.

From the paper
“the on-chip flow cytometer can then be designed as a GMP-compliant, closed-loop, disposable, multichannel device”

The conclusion names the cell therapy manufacturing setting (GMP-compliant, closed-loop, disposable device) in which these findings would be put into practice.

Pathway

Cell therapy manufacturers and instrument developers would integrate the chip with sorting modules into parallelized, closed-loop GMP cell processing devices.

Barrier

The chip still needs integration of a sorter, detectors and electronics, plus improved sensitivity and Z-focusing, before it can serve as a GMP-compliant manufacturing instrument.

Evaluated 25 Nov 2026 · GBD 2023 · glm_v1_3_3 · impact_profile_v1_3_3

Report an issue →

AI-generated, human-governed. Something look off? Contact us to request a review.

Keep reading

Related evaluations

Other papers Nabu has evaluated blind, starting with the same field.

  1. Flexible, Wearable and Stretchable Photonics: feature issue introduction

    Li, Guo, Song, et al. · 2026 · Physical sciences

    Primary claim

    Integrating photonics with dynamic, non-planar, and biological environments necessitates a departure from traditional rigid platforms toward systems that are inherently flexible, stretchable, and compliant

    Tentative
    Evidence Score
    2.9±0.1
    Reliability
    No concerns
  2. Non-contact photoacoustic imaging with a silicon photonics-based Laser Doppler Vibrometer

    Dieussaert, Baets, Jans, et al. · 2024 · Medical engineering

    Primary claim

    By scanning the LDV beam over the surface of the [silicone sample], 2D photoacoustic images were reconstructed of the sample.

    Supported
    Evidence Score
    3.6±0.1
    Reliability
    Caveats
  3. An Integrated Photonic Biosensing Platform for Pathogen Detection in Aquaculture

    Knoben, Graf, Borutta, et al. · 2024 · Medical engineering

    Primary claim

    The average net DNA binding signal [on the hybrid PIC platform] for 100 nM A. sal [single-stranded target DNA] is 1107 ± 85 pm

    Supported
    Evidence Score
    3.2±0.1
    Reliability
    Caveats

How the rubric works →