#reportingSource identity and reproducibility exceed demonstrated evidence
Common-wavelength operation establishes compatibility, not complete source indistinguishability. Different SHG spectra and fitted thresholds of 23 ± 4 W and 15 ± 3 W, together with evidence from only one device pair, also limit the reproducible-building-block inference.
↳ Abstract; Section III, Fig. 2c–d; Section IV.B; Conclusion
#methodological rigourEntanglement uncertainty excludes calibration and repeatability
The ±0.005 uncertainty comes from temporal blocks within one acquisition, not independent experimental runs. Systematic shot-noise calibration uncertainty is not quantified against the 0.13 margin below the separability bound.
↳ Section IV.C; Fig. 3d
#methodological rigourExtrema selection limits pump curve interpretation
Each pump-dependence point selects minimum squeezing and maximum antisqueezing from sequential measurements, which can favor extreme values in the fitted curves. The entanglement result instead averages across blocks and is not subject to this selection procedure.
↳ Sections IV.B–C; Fig. 3c–d