#methodological rigourIndependent loss calibration remains to be demonstrated
No cutback, resonator or known-reference measurement validates the conversion from scattered-light decay to guided-power loss. A good exponential fit alone cannot establish measurement accuracy.
↳ Section 2, Figure 2; Conclusion
#methodological rigourReplication and fit uncertainty remain unspecified
The paper does not state how many waveguides, chips or repeated measurements underlie each estimate, or provide uncertainty bounds. The analyzed physical length and quantitative background-adjustment procedure are also omitted.
↳ Section 2, Figures 2–4
#reportingMechanistic wording exceeds the measured width trend
The abstract attributes higher attenuation to scattering and reduced confinement without separately measuring those contributions. Section 2 also invokes bend leakage, leaving the mechanism unresolved.
↳ Abstract; Section 2, Figures 3–4