# Pulse oximetry screening for critical congenital heart disease in Tanzanian newborns: Diagnostic accuracy, sensitivity, and specificity in a low-resource healthcare setting.

- **DOI:** [10.1371/journal.pgph.0004904](https://doi.org/10.1371/journal.pgph.0004904)
- **Rating:** Limited (2.8 out of 5)
- **Verdict:** A large Tanzanian newborn cohort adds useful local screening evidence, but incomplete diagnostic verification limits its accuracy estimates.
- **Framework version:** v5.10
- **Pipeline version:** v0.4
- **Evaluation date:** 2026-10-23
- **Evaluation page:** https://www.nabu.science/paper/10.1371%2Fjournal.pgph.0004904

## Key claims

1. POX demonstrated a sensitivity of 50.0% (95% CI: 32.1-67.9), a specificity of 99.5% (95% CI: 99.4-99.7), a false-positive rate of 0.4%, and an overall accuracy of 99.5% (95% CI: 99.2-99.5)
   - Support: Tentative
   - Type: Primary finding
   - Basis: differential verification and incomplete follow-up leave false-negative ascertainment uncertain
   - Location: Abstract
2. Screenings conducted between 48 and 72 hours exhibited the highest diagnostic performance, AUC 0.79 (95% CI: 0.64-0.93), with a significant odds ratio (OR) of 5.31 (95% CI: 2.45-11.49, p = 0.00001)
   - Support: Tentative
   - Type: Secondary
   - Basis: few CCHD events and workflow-related timing confounding limit inference about optimal screening time
   - Location: Abstract; Results
3. Newborns with a birth weight <2.5 kg were less likely to have CCHD detected by POX, OR 0.403 (95% CI: 0.19-0.87, p = 0.021)
   - Support: Tentative
   - Type: Secondary
   - Basis: the small-event observational subgroup analysis does not establish a reliable basis for birth-weight-specific protocols
   - Location: Abstract; Results
4. Screening timing, delivery method, and birth weight influenced diagnostic performance.
   - Support: Tentative
   - Type: Secondary
   - Basis: delivery method was nonsignificant at p=0.234, while the remaining subgroup findings rest on few cases
   - Location: Conclusion
5. 74.5% of cases flagged by POX in our study required urgent medical attention for non-cardiac issues, highlighting POX’s broader potential as a screening tool for serious unrelated conditions.
   - Support: Tentative
   - Type: Secondary
   - Basis: the printed breakdown gives 21/51, or 41.2%, rather than 74.5%, without comparative evidence of broader screening benefit
   - Location: Discussion, summary paragraph

## Strengths

- **Large prospective cohort addresses local evidence** The study screened 10,630 newborns and followed screen-negative infants for six months, adding setting-specific evidence from an under-researched region.
  - Dimension: contribution
  - Location: Abstract; Results, Demographic characteristics and Echocardiography evaluation results
- **Systematic recruitment and fixed screening protocol** Screening reached 89% of eligible newborns and used a specified device and prespecified oxygen-saturation thresholds. Positive screens were assessed by paediatric cardiologists.
  - Dimension: methodological rigour
  - Location: Methods, Enrollment and Echocardiography evaluation; Results, Demographic characteristics
- **Comparisons span contrasting clinical resource settings** The Discussion compares the observed sensitivity with high-income benchmarks and other low-resource studies, while identifying limits for lower-level facilities and high-altitude regions.
  - Dimension: positioning
  - Location: Discussion; Strengths and limitations

## Limitations

- **Unequal verification leaves missed cases uncertain** Negative screens were not routinely verified by echocardiography, and 13.3% lacked follow-up. Two positive-screen newborns who died before echocardiography were excluded, further limiting the accuracy analysis.
  - Dimension: methodological rigour
  - Location: Methods, Echocardiography evaluation; Results, POX screening results and Echocardiography evaluation results
- **Delivery conclusion conflicts with reported analysis** The Conclusion says delivery method influenced performance, but the Results report OR=1.613, 95% CI 0.73–3.54 and p=0.234. The analysis does not establish the stated influence.
  - Dimension: reporting
  - Location: Results, Accuracy of the pulse oximetry test and the associated factors; Conclusion
- **Subgroup evidence cannot establish tailored protocols** Timing and birth-weight analyses rest on only 30 confirmed CCHD cases, with timing also affected by hospital workflow and discharge. These findings provide questions for validation rather than established grounds for protocol tailoring.
  - Dimension: methodological rigour
  - Location: Results, POX screening results and associated factors; Discussion

## Reliability checks

Concerns (4 of 4 checks).

- **Results check:** Concern
  - Screening timing, delivery method, and birth weight influenced diagnostic performance: "Screening timing, delivery method, and birth weight influenced diagnostic performance." (Conclusion) vs "it did not significantly impact the accuracy (p = 0.234; S2 Appendix and S3 Appendix)." (Results, Accuracy of the pulse oximetry test and the associated factors). The conclusion asserts a delivery-method influence although the Results explicitly report that association as nonsignificant. → Not supported by the paper's own data
  - 74.5% of cases flagged by POX in our study required urgent medical attention for non-cardiac issues, highlighting POX's broader potential: "74.5% of cases flagged by POX in our study required urgent medical attention for non-cardiac issues" (Discussion, summary paragraph) vs "Of the 34 false positives (0.45%), 8 (23.5%) were normal, while 26 (76.5%) had other conditions requiring urgent medical intervention" (Results, Echocardiography evaluation results). The printed breakdown gives 10 respiratory plus 11 infectious cases, so urgent non-cardiac conditions account for 21/51=41.2% of all flagged newborns, not 74.5%. → Changes the size materially
  - POX demonstrated a sensitivity of 50.0% (95% CI: 32.1-67.9), a specificity of 99.5% (95% CI: 99.4-99.7): "the specificity estimate was highly precise (98.8% to 99.2%)" (Strengths and limitations) vs "a specificity of 99.5%" (Abstract). The Limitations interval excludes the reported point estimate, while the printed counts give specificity 9,155/(9,155+34)=99.63%. → Conclusion holds
  - Minor reporting notes (5):
    - Descriptive: "18 (35.3%) had congenital heart disease (CHD), and 15 (83.3%) were classified as critical" (Abstract) vs "non-critical congenital heart defects (5; 19.2%)" (Results, Echocardiography evaluation results). The abstract implies 18−15=3 non-critical CHD cases, whereas the Results report five. → Conclusion holds
    - Descriptive: "With a follow-up rate of 86.7% (9,170/10,574)" (Abstract) vs "Among the 10,579 with negative POX screen results" (Results, Echocardiography evaluation results). 10,630 screened minus 51 positives equals 10,579 negatives, and 9,170/10,579 rounds to 86.7%. → Conclusion holds
    - Descriptive: "The diagnostic odds ratio (DOR) was 221.5 (100.4-488.6)" (Results, Accuracy of the pulse oximetry test and the associated factors) vs "There were 34 false positives and 9,155 true negatives" (Results, Accuracy of the pulse oximetry test and the associated factors). The printed contingency counts give DOR=(15×9,155)/(15×34)=269.26 rather than 221.5. → Conclusion holds
    - Descriptive: "45.0% (95% CI; 66.8-88.6) for cono-truncal lesions" (Results, Accuracy of the pulse oximetry test and the associated factors). The reported 66.8–88.6% confidence interval does not contain its 45.0% point estimate. → Effect on result unclear
    - Descriptive: "caesarean section deliveries contributed to 5,721 (58.5%)" (Results, Demographic characteristics of screened newborns) vs "Males contributed to 5,721 newborns (54.0%)" (Results, Demographic characteristics of screened newborns). The same count is assigned different percentages, and 5,721/10,630=53.82%, not 58.5%. → Effect on result unclear
- **Reporting and transparency:** Noted. The Methods and Limitations assign incompatible missing-data mechanisms to follow-up losses. Ethics approval is declared, but a published protocol alone does not establish prospective registration.
  - Conflict of interest: not stated · Ethics approval: declared
- **Reference integrity:** Clean
  - 35 references in manuscript
  - 35 of 35 checkable references found in an index
  - References not verified automatically are resolved manually within 24 hours.

## Post-publication record

- **Status:** Clean
- **Summary:** No retraction or expression of concern found in Crossref (includes Retraction Watch).
- **Last checked:** 2026-10-23