# Demonstration of temperature monitoring of the curing process in resin transfer moulding using integrated phase-shifted Bragg grating sensors

- **DOI:** [10.1364/optcon.522327](https://doi.org/10.1364/optcon.522327)
- **Rating:** Limited (2.8 out of 5)
- **Verdict:** An integrated photonic probe tracks an epoxy-curing cycle to 120 °C, but non-co-located reference measurements leave its claimed accuracy unvalidated.
- **Framework version:** v5.13
- **Pipeline version:** v0.4
- **Evaluation date:** 2026-11-25
- **Evaluation page:** https://www.nabu.science/paper/10.1364%2Foptcon.522327

## Key claims

1. The embedded phase-shifted Bragg grating sensors exhibit high sensitivity (75 pm/°C) and linearity (R2=0.995), operating up to 120 °C, allowing precise temperature measurements throughout the curing cycle.
   - Support: Tentative
   - Type: Primary finding
   - Basis: single-probe testing with calibration only through 100 °C and no co-located curing reference
   - Location: Conclusions
2. The experimental results show consistency with previously reported simulations, throughout the calibration, dry run characterization and the curing process, up to 120 °C, allowing for accurate monitoring of the curing process.
   - Support: Tentative
   - Type: Secondary
   - Basis: single-probe comparison with measured sensitivity below simulation and no independent validation of interface accuracy
   - Location: Abstract
3. The optical sensors went through calibration, dry run characterization, and curing process measurements, with the small variations in measurement caused by the local temperature profile in the tool-resin interface.
   - Support: Tentative
   - Type: Secondary
   - Basis: differently located sensors cannot identify the cause of their measurement differences
   - Location: Conclusions

## Strengths

- **Photonic probe tested during actual curing** A packaged silicon-photonic sensor records reflection spectra during resin infusion and subsequent heating to 120 °C. This extends the earlier dry-run work to an operating composite tool.
  - Dimension: contribution
  - Location: Introduction; Section 3, Figs. 11–13
- **Calibration supports a consistent temperature response** The dry-run calibration is referenced to thermocouples through 100 °C. Reported wavelength shifts of 1.875 nm over 25 °C and 4.125 nm over 55 °C both match the stated 75 pm/°C sensitivity.
  - Dimension: methodological rigour
  - Location: Section 3, first paragraph; Figs. 8–9
- **Probe failure and spectral differences disclosed** The results report loss of the second probe's signal above its 30–40 °C test range and describe the broader-than-simulated reflection dip. These disclosures expose relevant packaging and fabrication constraints.
  - Dimension: positioning
  - Location: Section 3, spectral-bandwidth discussion and Fig. 10

## Limitations

- **Reference placement prevents independent accuracy validation** Thermocouples near the heaters and an optical sensor at the tool edge show a 1–5% difference. That comparison cannot separate sensor error from a real thermal gradient or establish location as the cause of the difference.
  - Dimension: methodological rigour
  - Location: Methods §2.3, Fig. 7; Section 3, Fig. 12
- **Single surviving probe limits operating conclusions** Only one probe is followed through a reported curing cycle after the second loses signal during dry testing. Repeatability, hysteresis and drift are not established, and the conclusion does not trace the disclosed probe failure into the scope of its precision claim.
  - Dimension: positioning
  - Location: Section 3, Figs. 10–13; Conclusions
- **Headline precision exceeds the reported validation** The headline attaches R2=0.995 to operation through 120 °C, although dry calibration stops at 100 °C and the curing-stage fit is R2=0.988. The Abstract also names RTM-6 while Methods identifies Epikote RIMR 035c/Epikure RIMH 037, leaving the tested material uncertain.
  - Dimension: reporting
  - Location: Abstract; Methods §2.3; Section 3, Fig. 13; Conclusions

## Reliability checks

Caveats (4 of 4 checks).

- **Results check:** Noted
  - "linearity (R2=0.995) up to 120 °C, enabling precise temperature measurements" (Abstract, sensitivity and linearity statement) vs "The sensor was calibrated up to 100°C using the reference thermocouple probe" (Section 3, first paragraph). Compare the headline range with the stated calibration endpoint and the separately reported dry-calibration and curing-stage fits in Section 3 and Fig. 13. The result itself is unaffected.
- **Reporting and transparency:** Noted. The Abstract and Methods identify different resin systems, leaving the demonstrated material uncertain. Conflict-of-interest and data-availability statements are present.
  - Conflict of interest: declared · Data availability: declared
- **Reference integrity:** Clean. 30 of 30 checkable references verified
  - 30 references in manuscript
  - 11 references confirmed by manual review
  - References not verified automatically are resolved manually within 24 hours.

## Post-publication record

- **Status:** Clean
- **Summary:** No retraction or expression of concern found in Crossref (includes Retraction Watch).
- **Last checked:** 2026-11-25