#methodological rigourFar-field measurement plane lacks adequate validation
The lensless camera is positioned 40 cm from the 2 mm aperture, versus a conventional far-field distance of approximately 5 m. Coordinate conversion alone does not establish the reported width as far-field divergence, and calibration and measurement uncertainty are not documented.
↳ Section 3; Section 4.2
#reportingWavelength labels reverse between text and caption
Section 4.2 assigns Fig. 6(a) and (b) to 1568 nm and 1497 nm, while the caption reverses them. The minimum reported performance values survive, but their wavelength-specific attribution remains unresolved.
↳ Section 4.2, first paragraph; Fig. 6 caption
#methodological rigourPower collection and transverse validation remain incomplete
The collection and calibration method behind the reported surface-emitted power above 1.1 mW is unspecified. The simulated 10.63° transverse beam width is not experimentally verified, limiting validation of the complete line-shaped output.
↳ Section 3; Section 4.1; Section 4.2; Fig. 3(c)