# Sensing of ceftazidime antibiotics in human plasma through waveguide-enhanced Raman spectroscopy assisted by a mesoporous layer

- **DOI:** [10.1063/5.0278252](https://doi.org/10.1063/5.0278252)
- **Rating:** Sound (3.2 out of 5)
- **Verdict:** Mesoporous waveguide Raman sensing extends ceftazidime measurement into spiked plasma, but incomplete analytical validation limits its sensitivity and clinical claims.
- **Framework version:** v5.13
- **Pipeline version:** v0.4
- **Evaluation date:** 2026-11-25
- **Evaluation page:** https://www.nabu.science/paper/10.1063%2F5.0278252

## Key claims

1. Leveraging waveguiding signal enhancement, chemical enrichment, and molecular sieving, we demonstrate detection of 1 mg/l ceftazidime in plasma within 2 min.
   - Support: Tentative
   - Type: Primary finding
   - Basis: single-chip technical repeats without a matched blank-plasma detection threshold under the two-minute acquisition
   - Location: Abstract
2. Ceftazidime quantification is achieved in plasma over 0, 20, 40, and 60 mg/l, demonstrating its potential for clinical use.
   - Support: Supported
   - Type: Secondary
   - Basis: independent-chip calibration in spiked pooled plasma without independent accuracy or patient-matrix validation
   - Location: Abstract; Fig. 4
3. Using the functionalized waveguide platform, we investigated β-lactam antibiotics spiked into human plasma. Distinct Raman spectral features were consistently observed at clinically relevant concentrations, indicating the sensor’s broad applicability across multiple antibiotic compounds.
   - Support: Tentative
   - Type: Secondary
   - Basis: testing ceftazidime alone cannot establish performance across multiple antibiotic compounds
   - Location: Conclusion, p. 5, paragraph 4
4. Notably, when measuring a concentration of 60 mg/L of antibiotics using both WERS and free-space configurations, under identical laser power and integration time, the waveguide-based signal will be ∼110 times stronger than that of the free-space setup
   - Support: Tentative
   - Type: Secondary
   - Basis: the free-space comparator at 60 mg/L is not explained despite unresolved background at 200 mg/l
   - Location: Results, p. 4, paragraph 4

## Strengths

- **Waveguide sensing reaches a complex plasma matrix** The study moves functionalized waveguide Raman sensing beyond simple solutions to ceftazidime-spiked pooled human plasma. The resulting concentration response establishes a bounded proof of concept for this matrix.
  - Dimension: contribution
  - Location: Introduction; Results, Figs. 2–4
- **Independent chips support the calibration experiment** Each of the 12 calibration samples is measured on a new chip, with three samples per concentration. Normalization to the coating's 1600 cm⁻¹ peak addresses variation in the optical platform.
  - Dimension: methodological rigour
  - Location: Results, Fig. 4
- **Spectral controls support assignment of ceftazidime bands** Water, blank plasma, and an uncoated waveguide provide background comparisons. A concentrated free-space ceftazidime reference supports assignment of the observed Raman features.
  - Dimension: methodological rigour
  - Location: Results, Figs. 2–3

## Limitations

- **Low concentration detection lacks independent validation** The 1 mg/l result uses three waveguides on one chip exposed to one sample. No defined blank-based detection threshold or matched blank-plasma noise characterization is reported for the two-minute, smoothed acquisition.
  - Dimension: methodological rigour
  - Location: Results, Fig. 5
- **Calibration does not establish independent assay accuracy** The 12 reported samples comprise the calibration levels, without independent quality-control samples or a validation set. The 17.0% RSD at 20 mg/l also exceeds the paper's cited typical 15% criterion.
  - Dimension: methodological rigour
  - Location: Results, Fig. 4 and reproducibility paragraph
- **Conclusion extends beyond the single tested antibiotic** The Conclusion asserts broad applicability across multiple antibiotic compounds, although the experiments test only ceftazidime. The demonstrated scope is therefore narrower than this central statement suggests.
  - Dimension: reporting
  - Location: Conclusion

## Reliability checks

Caveats (4 of 4 checks).

- **Results check:** Noted
  - "the Raman signal from 200 mg/l of ceftazidime cannot be readily separated from the solution background" (Results, free-space reference paragraph (Fig. 3)) vs "than that of the free-space setup" (Results, 110-fold enhancement statement). The stated inability to separate the free-space signal at 200 mg/l from background leaves the derivation of the free-space comparator at 60 mg/L unexplained; no corrected enhancement ratio can be calculated. The effect on the result is unclear.
  - Minor reporting notes (1):
    - "We report the WERS detection of ceftazidime, a model antibiotic, at therapeutically relevant concentrations ranging from 8 to 80 mg/l." (Introduction, p. 2, paragraph 6) vs "a total of 12 spiked plasma samples are prepared, each with 0, 20, 40, or 60 mg/l ceftazidime (3 repeats each)." (Results, p. 4, paragraph 5). The introductory range of 8–80 mg/l differs from the listed calibration concentrations of 0, 20, 40, and 60 mg/l. The result itself is unaffected.
- **Reporting and transparency:** Clean. Commercial pooled plasma was used, conflicts of interest are disclosed, and supporting data are offered on reasonable request. No text-grounded research-conduct finding is established.
  - Conflict of interest: declared · Data availability: declared
- **Reference integrity:** Clean. 33 of 34 checkable references verified
  - 37 references in manuscript
  - 3 have no canonical index record — counted, but not index-checkable
  - 3 references confirmed by manual review
  - References not verified automatically are resolved manually within 24 hours.

## Post-publication record

- **Status:** Clean
- **Summary:** No retraction or expression of concern found in Crossref (includes Retraction Watch).
- **Last checked:** 2026-11-25