# High-performance photon number resolving detectors for 850–950 nm wavelength range

- **DOI:** [10.1063/5.0204340](https://doi.org/10.1063/5.0204340)
- **Rating:** Limited (2.9 out of 5)
- **Verdict:** NbTiN detectors extend rising-edge photon-number resolution to seven peaks, but high efficiency and maximum resolution are demonstrated on different devices and wavelengths.
- **Framework version:** v5.13
- **Pipeline version:** v0.4
- **Evaluation date:** 2026-11-25
- **Evaluation page:** https://www.nabu.science/paper/10.1063%2F5.0204340

## Key claims

1. In Fig.4a and Fig.4b multiple photon events can be resolved, from 3 to 7 photons corresponding to the used photon fluxes of 350 kcps and 900 kcps respectively [at 1064 nm].
   - Support: Tentative
   - Type: Primary finding
   - Basis: photon-number labels checked against a Poisson distribution whose mean is derived from those same assigned peaks
   - Location: Section 5, paragraph 2
2. The averaged SDE over three measurements, shown in fig.5a, was 94.5%.
   - Support: Supported
   - Type: Secondary
   - Basis: reference-arm metrology supports the measured average for this detector, with limited replication and incompletely specified uncertainty propagation
   - Location: Section 6
3. In this paper, we demonstrate NbTiN-based SNSPDs with > 94% system detection efficiency, sub-11 ps timing jitter for one photon, and sub-7 ps for 2-photon.
   - Support: Tentative
   - Type: Secondary
   - Basis: the presented characterization does not establish the claimed fast-jitter performance alongside high efficiency
   - Location: Abstract

## Strengths

- **Seven resolved peaks with conventional readout** Rising-edge timing histograms distinguish up to seven photon-response peaks using a conventional cryogenic amplifier. This extends the resolved-number range beyond the earlier demonstrations cited in the introduction.
  - Dimension: contribution
  - Location: Introduction; Section 5, Figures 3–4
- **Traceable efficiency measurements with repeated acquisition** The 940 nm efficiency measurement uses a reference arm and a NIST-traceable power meter. Three measurements yield a reported average system detection efficiency of 94.5%.
  - Dimension: methodological rigour
  - Location: Section 6, Figure 5a
- **Application motivation grounded in relevant wavelengths** The introduction explains the relevance of 850–950 nm operation to semiconductor quantum dots and cesium-based quantum memories. It also places the detector concept alongside earlier SNSPD photon-number-resolution work.
  - Dimension: positioning
  - Location: Introduction, references 29–33

## Limitations

- **Headline specifications combine separately demonstrated capabilities** Seven-photon resolution is measured at 1064 nm on a different device from the detector achieving 94.5% efficiency at 940 nm. The abstract's sub-11 ps and sub-7 ps jitter figures also lack corresponding reported results, while Section 6 reports 18.4 ps for the high-efficiency device.
  - Dimension: reporting
  - Location: Title and Abstract; Sections 5–7
- **Photon labels receive only conditional validation** The Poisson prediction uses a mean calculated from the same assigned photon peaks, providing a consistency check rather than independent photon-number ground truth. PNR is not characterized within the target 850–950 nm band, and per-peak classification fidelity is not quantified.
  - Dimension: methodological rigour
  - Location: Sections 4–5, Figure 4
- **Proposed improvements lack measured experimental comparison** The theoretical discussion identifies readout noise, bandwidth and detector parameters as routes to better PNR performance. It is not quantitatively tested against the measured detector results.
  - Dimension: methodological rigour
  - Location: Section 3, Figure 2; Sections 5–6

## Reliability checks

Caveats (4 of 4 checks).

- **Results check:** Noted
  - "giving an error of 1.3% in the system detection efficiency" (Section 6, SDE uncertainty) vs "The measurement uncertainty can be inferred from the power-meter uncertainty (1.1%)" (Section 6). Assuming independent quadrature, the stated 1.1% and 0.5% components combine as sqrt(1.1² + 0.5²) = approximately 1.21%, rather than 1.3%, without changing the efficiency point estimate. The result itself is unaffected.
  - "it is clear that the distribution for photon number zero is overestimated while the peaks for photon numbers of 1 and 2 are slightly underestimated" (Section 5, paragraph 1) vs "The measured photon number zero is, contrary to expectations, slightly lower than the expected value which we ascribe to the low extinction ratio" (Figure 4 caption). The main text does not specify whether 'overestimated' describes the prediction or measurement, so comparison with the caption establishes ambiguous wording but not an unambiguous reversal. The effect on the result is unclear.
- **Reporting and transparency:** Noted. Section 3.2 cites Figure 4b for a domain-lifetime relationship, whereas that figure presents PNR measurements. Data availability on request is declared and is not treated as suspicious.
  - Data availability: declared
- **Reference integrity:** Clean. 36 of 37 checkable references verified
  - 37 references in manuscript
  - 20 references confirmed by manual review
  - References not verified automatically are resolved manually within 24 hours.

## Post-publication record

- **Status:** Clean
- **Summary:** No retraction or expression of concern found in Crossref (includes Retraction Watch).
- **Last checked:** 2026-11-25