# Non-contact photoacoustic imaging with a silicon photonics-based Laser Doppler Vibrometer.

- **DOI:** [10.1038/s41598-024-74266-y](https://doi.org/10.1038/s41598-024-74266-y)
- **Rating:** Sound (3.6 out of 5)
- **Verdict:** Silicon-photonics vibrometry enables contactless photoacoustic phantom imaging, but idealized reflective samples limit the demonstration's biomedical reach.
- **Framework version:** v5.13
- **Pipeline version:** v0.4
- **Evaluation date:** 2026-11-25
- **Evaluation page:** https://www.nabu.science/paper/10.1038%2Fs41598-024-74266-y

## Key claims

1. By scanning the LDV beam over the surface of the [silicone sample], 2D photoacoustic images were reconstructed of the sample.
   - Support: Supported
   - Type: Primary finding
   - Basis: phantom reconstructions demonstrate feasibility despite limited quantitative validation of image performance
   - Location: Abstract
2. For [clear, flat PDMS-based samples with an ink solution-filled channel embedded], the photonics-based detection system presented a detection limit on par with or even better than commercial systems, with a bandwidth of up to around 3.5 MHz.
   - Support: Supported
   - Type: Secondary
   - Basis: commercial-reference calibration supports the prepared-sample comparison despite incompletely reported matching conditions and repeatability
   - Location: Conclusion and Outlook
3. We detect small photoacoustic signals for ink concentrations down to 0.01 % corresponding to an absorption coefficient of around 1.2 cm-1.
   - Support: Supported
   - Type: Secondary
   - Basis: the phantom concentration series supports detection without independent repeats or a quantified detection uncertainty
   - Location: Results and Discussion, Non-contact Photoacoustic Imaging with Silicon Photonics-based LDV
4. Reconstruction images at different depths (Fig. 3e) demonstrate the capability to determine the position and depth of the channel.
   - Support: Supported
   - Type: Secondary
   - Basis: multiple channel depths are reconstructed without a quantified localization-error analysis
   - Location: Results, Non-contact photoacoustic imaging; Fig. 3e

## Strengths

- **Commercial reference anchors the detector characterization** Noise spectra and pulse responses are compared with a commercial Polytec LDV. Pre-measurement ellipse fitting addresses IQ imbalance before demodulation.
  - Dimension: methodological rigour
  - Location: Methods, Performance Characterization Methods; Fig. 2a–e
- **Integrated hardware demonstrates contactless phantom imaging** An on-chip LDV paired with a compact pulsed laser diode reconstructs ink-filled channels in silicone. Depth and concentration series extend the demonstration beyond a single recorded waveform.
  - Dimension: contribution
  - Location: Results and Discussion; Figs. 3 and 5
- **Translation limits are linked to consequences** The discussion connects surface scattering, physiological absorption, averaging time and electrical interconnects to sensitivity, imaging depth, motion and array scaling. Prospective biomedical performance is distinguished from the demonstrated phantom results.
  - Dimension: positioning
  - Location: Results and Discussion; Conclusion and Outlook

## Limitations

- **Detection uncertainty lacks independent repeat characterization** The concentration-series SNR plot gives no independent repeat counts or error bars, and its RMSE denominator is undefined. Pulse averaging does not establish variability across repeated fills, alignments or phantom preparations.
  - Dimension: methodological rigour
  - Location: Results, concentration series; Fig. 5e
- **Prepared surfaces restrict the demonstrated applicability** Detection uses clear, glass-molded PDMS with a specular surface and opposite-side illumination. These conditions do not establish performance on rough, scattering biological samples.
  - Dimension: methodological rigour
  - Location: Methods, Silicone Sample; Fig. 1a; Conclusion and Outlook
- **Image performance remains only partly quantified** The approximately 230 µm resolution is theoretical rather than measured, and channel-localization error is not quantified. Realignment and illumination differences also confound amplitude comparisons between depths.
  - Dimension: methodological rigour
  - Location: Results, Non-contact Photoacoustic Imaging; Fig. 3e

## Reliability checks

Caveats (4 of 4 checks).

- **Results check:** Clean. The reported numbers and statistics are consistent with the stated results.
- **Reporting and transparency:** Noted. Methods misidentify the sensitivity panel as Fig. 2d rather than Fig. 2e, without changing the result. Data and signal-processing code availability and competing interests are declared.
  - Conflict of interest: declared · Data availability: declared · Code availability: declared
- **Reference integrity:** Clean
  - 37 references in manuscript
  - 36 of 36 checkable references found in an index
  - 1 are books, websites or datasets — counted, but not index-checkable
  - References not verified automatically are resolved manually within 24 hours.

## Post-publication record

- **Status:** Clean
- **Summary:** No retraction or expression of concern found in Crossref (includes Retraction Watch).
- **Last checked:** 2026-11-25