# On-chip flow cytometer using integrated photonics for the detection of human leukocytes.

- **DOI:** [10.1038/s41598-024-60708-0](https://doi.org/10.1038/s41598-024-60708-0)
- **Rating:** Sound (3.7 out of 5)
- **Verdict:** Integrated photonics enables lymphocyte–monocyte discrimination against a commercial benchmark, but single-donor testing confines the result to a proof of concept.
- **Framework version:** v5.13
- **Pipeline version:** v0.4
- **Evaluation date:** 2026-11-25
- **Evaluation page:** https://www.nabu.science/paper/10.1038%2Fs41598-024-60708-0

## Key claims

1. The optical device is capable of detecting the side- and forward-scattered light of white blood cells, with sufficient resolution for the discrimination of monocytes and lymphocyte populations in a peripheral blood mononuclear cell (PBMC) sample.
   - Support: Supported
   - Type: Primary finding
   - Basis: complementary benchmark and cell-subset measurements in a single donor
   - Location: Introduction, final device-capability paragraph
2. the composition of the cell population as found on-chip, 15.7% monocytes, and 77.4% lymphocytes, corresponds well to the composition given by the benchmark flow cytometer, 22.6% and 75.9%.
   - Support: Tentative
   - Type: Secondary
   - Basis: single-donor comparison with a 6.9-percentage-point monocyte difference and no quantified agreement uncertainty
   - Location: Results, Detection of full PBMC populations; Fig. 6
3. On-chip, 95.3% of the CD3 + cells within Q1 fall within the lymphocyte gate, compared to 94.5% on the benchmark flow cytometer.
   - Support: Supported
   - Type: Secondary
   - Basis: single-donor back-gating restricted to detected CD3-positive cells
   - Location: Results and Discussion, Fluorescence, paragraph 1
4. the on-chip discrimination factor of FSC in ALL mode amounts to 1.4, which is on par with the discriminating factor of 1.4 obtained on the benchmark flow cytometer.
   - Support: Supported
   - Type: Secondary
   - Basis: matching measured discrimination factors in isolated populations from one donor
   - Location: Results, Detection of isolated PBMC populations

## Strengths

- **Integrated scatter collection distinguishes leukocyte populations** Monolithically integrated illumination and collection optics resolve lymphocyte and monocyte populations in the tested PBMC sample. This extends the bead-discrimination and counting capabilities described for earlier integrated devices.
  - Dimension: contribution
  - Location: Introduction; Results, Figs. 5–6
- **Parallel benchmarking anchors the device measurements** Every sample was measured on a commercial CytoFLEX S instrument. Calibration beads, isolated leukocyte populations and CD3 fluorescence back-gating provide complementary checks on the on-chip signals.
  - Dimension: methodological rigour
  - Location: Methods, Benchmark cytometry measurements; Results, Figs. 3 and 5–7
- **Physical limitations connect directly to performance** The analysis links limited collection efficiency and particle-height variation to sensitivity and population spread. Two interrogation positions test the proposed contribution of inertial focusing to bimodal bead side-scatter signals.
  - Dimension: positioning
  - Location: Results, Fig. 4; Conclusions and perspectives

## Limitations

- **Single donor limits biological performance validation** All PBMC experiments use aliquots from one donor, and three repeated measurements do not establish donor-to-donor performance. Chip-to-chip reproducibility is also unquantified.
  - Dimension: methodological rigour
  - Location: Methods, Cell preparation; Results, Detection of full PBMC populations
- **Monocyte agreement wording exceeds the comparison** The on-chip monocyte fraction is 15.7%, compared with 22.6% on the benchmark: a 6.9-percentage-point difference, approximately 31% relative. Describing these compositions as corresponding well understates the unresolved difference.
  - Dimension: reporting
  - Location: Results, Detection of full PBMC populations; Fig. 6
- **Fluorescence checks only part of classification** CD3-positive back-gating checks a subset of lymphocytes rather than all lymphocytes or monocyte identity. High fluorescence background and zero assignment to undetected fluorescence events further limit this validation.
  - Dimension: methodological rigour
  - Location: Results, Fluorescence; Fig. 7

## Reliability checks

Caveats (4 of 4 checks).

- **Results check:** Clean. The reported numbers and statistics are consistent with the stated results.
- **Reporting and transparency:** Noted. The supplied text lacks an ethics, waiver or consent declaration for the human biobank material and contains an incorrect Figure 7 panel reference. Commercial funding, competing interests and request-based data access are disclosed.
  - Conflict of interest: declared · Ethics approval: not stated · Data availability: declared
- **Reference integrity:** Clean
  - 29 references in manuscript
  - 29 of 29 checkable references found in an index
  - Citation details diverge for 2 references:
    - #6 Neurauter, A. A. et al. Cell isolation and expansion using dynabeads ®. In Cell Separation 41–73. (Springer, Berlin). 10.1007/10_2007_072. 10.1007/10_2007_072 17680228 (DOI resolves to: “Cell Isolation and Expansion Using Dynabeads ®” — Neurauter (Advances in Biochemical Engineering/Biotechnology). Differs from the DOI’s record on: container) https://doi.org/10.1007/10_2007_072
    - #22 Verellen, N. et al. Integrated photonics for miniature flow cytometry. in (2019). 10.7567/ssdm.2017.f-4-01. (DOI resolves to: “Integrated photonics for miniature flow cytometry” — Verellen, 2017 (Extended Abstracts of the 2017 International Conference on Solid State Devices and Materials). Differs from the DOI’s record on: year) https://doi.org/10.7567/ssdm.2017.f-4-01
  - References not verified automatically are resolved manually within 24 hours.

## Post-publication record

- **Status:** Clean
- **Summary:** No retraction or expression of concern found in Crossref (includes Retraction Watch).
- **Last checked:** 2026-11-25