# Flow Injection-Based Refractive Index Sensing with a Si3N4 Photonic Crystal Nanobeam-Microring Fano Resonator.

- **DOI:** [10.1021/acsaom.5c00500](https://doi.org/10.1021/acsaom.5c00500)
- **Rating:** Limited (2.7 out of 5)
- **Verdict:** Flow-injection testing establishes dual-readout Si3N4 sensing, but the comparison does not establish Fano asymmetry as the cause of improved intensity performance.
- **Framework version:** v5.13
- **Pipeline version:** v0.4
- **Evaluation date:** 2026-11-25
- **Evaluation page:** https://www.nabu.science/paper/10.1021%2Facsaom.5c00500

## Key claims

1. the PhCN-MRR exhibited comparable Δλ sensitivities (∼111 to 113 nm/RIU) but delivered enhanced ΔI responsivity and improved contrast at low concentrations (e.g., 2 mg/mL) due to its asymmetric resonance slopes.
   - Support: Tentative
   - Type: Primary finding
   - Basis: missing reference-ring intensity calibration and unmatched resonances prevent attribution of the claimed intensity advantage to asymmetry
   - Location: Abstract
2. Compared with a conventional microring resonator (MRR) featuring symmetric Lorentzian responses, the PhCN-MRR exhibited comparable Δλ sensitivities (∼111 to 113 nm/RIU)
   - Support: Supported
   - Type: Secondary
   - Basis: repeated injections support similar wavelength sensitivities for the selected resonances without establishing device-to-device variability
   - Location: Abstract; Results (Fig. 4c)
3. In contrast, ΔI stands for a reduced LOD and LOQ (3.5-fold) and an enhanced detection capability (28-fold) compared to Δλ.
   - Support: Tentative
   - Type: Secondary
   - Basis: correlated transient calibration points and an unspecified noise definition limit a performance comparison based partly on incompatible slope units
   - Location: Results, Transduction metrics (Table 1)
4. At concentrations above ∼5 mg/mL, the [PhCN-MRR] ΔI response is no longer linear, whereas Δλ remains strictly proportional to the RI change
   - Support: Supported
   - Type: Secondary
   - Basis: the tested concentration series directly shows intensity-response flattening while wavelength tracking retains linearity
   - Location: Results and Discussion, Dynamic Range Limitations, Figure 7

## Strengths

- **Dynamic flow testing extends device characterization** The Si3N4 nanobeam–microring platform is evaluated under repeated glucose flow injections with both wavelength and intensity readouts, extending its characterization beyond stopped-flow spectra.
  - Dimension: contribution
  - Location: Figures 4–6; Table 2
- **Fabrication and fluidic procedures are specified** The experimental section describes single-etch fabrication, PSA/PDMS flow-cell assembly and a defined injection protocol. Leak testing at 132 μL/min for 5.5 hours provides an operational check on the fluidic integration.
  - Dimension: methodological rigour
  - Location: Experimental Section, Device Design and Fabrication; Microfluidics Integration; Flow Injection Analysis Methodology
- **Readout trade-offs receive direct experimental attention** Figure 7 demonstrates intensity-response distortion near 5 mg/mL while wavelength tracking remains interpretable. The accompanying discussion identifies thermal, laser and coupling contributions to uncertainty.
  - Dimension: positioning
  - Location: Dynamic Range Limitations, Figure 7; Noise Sources Affecting Measurement Uncertainty and Error Mitigation

## Limitations

- **Intensity advantage lacks a matched baseline** Table 1 and Figure 5 calibrate intensity response only for the PhCN-MRR, leaving its claimed advantage over the conventional ring unquantified. Different selected resonances and transmission characteristics also prevent isolating asymmetry as the cause.
  - Dimension: methodological rigour
  - Location: Abstract; FIA Protocol and Sensing Workflow; Transduction Metrics, Figure 5 and Table 1
- **Detection metrics lack a reproducible comparison** The noise quantity used to calculate LOD and LOQ is unspecified, and substituting the printed residual SDs does not reproduce Table 1. The 28-fold claim compares slopes in incompatible units, while calibration uses related points along an FIA peak rather than independent concentration standards.
  - Dimension: methodological rigour
  - Location: Transduction Metrics, Figure 5e–f and Table 1
- **Conflicting dimensions obscure the tested geometry** The fabrication section specifies a ring radius near 16 μm and five nanobeam slots, but the later design discussion says 35 μm and ten slots were employed. This leaves the physical configuration underlying the reported results ambiguous.
  - Dimension: methodological rigour
  - Location: Device Design and Fabrication; Dynamic Range Limitations

## Reliability checks

Caveats (3 of 4 checks).

- **Results check:** Noted
  - "ΔI-right [a.u.] | 0.5055 a.u./(mg/mL) | 0.0184 au | 0.9794 | 0.0471 au | 9.6669 | 0.0238 | 0.0794" (Table 1, row 'ΔI-right') vs "Limit of detection (3σ)" (Table 1 footnote c). Assuming σ is the tabulated residual SD, 3×SD/slope gives LODs of 0.0616, 0.2411 and 0.2795 mg/mL for Δλ, ΔI-left and ΔI-right, but the paper does not explicitly identify the σ used. The effect on the result is unclear.
  - "ΔI stands for a reduced LOD and LOQ (3.5-fold)" (Results, Transduction metrics, text after Table 1) vs "ΔI-left [a.u.] | 0.3297 au/(mg/mL) | 0.0174 au | 0.9846 | 0.0265 au | 8.4891 | 0.0485 | 0.1618" (Table 1, row 'ΔI-left'). The printed LOD ratios are 0.0832/0.0238 = 3.50 for ΔI-right and 0.0832/0.0485 = 1.72 for ΔI-left, so the stated reduction holds for the right slope only. The result itself is unaffected.
  - "Both devices share a nominal ring radius of R MRR ≈ 16 μm" (Experimental Section, Device Design and Fabrication) vs "In our design, a larger ring radius (e.g., R MRR = 35 μm) was first employed to increase Q i" (Results and Discussion, Dynamic Range Limitations). The fabrication section specifies a nominal radius of approximately 16 μm, while the later description says a 35 μm radius was employed, leaving the tested geometry ambiguous. The effect on the result is unclear.
- **Reporting and transparency:** Noted. Figure 3's caption and accompanying text identify panel (a) as different device types. Competing interests are disclosed, and underlying data are offered upon reasonable request.
  - Conflict of interest: declared · Data availability: declared · Code availability: not stated
- **Reference integrity:** Clean
  - 50 references in manuscript
  - 50 of 50 checkable references found in an index
  - 1 reference confirmed by manual review
  - References not verified automatically are resolved manually within 24 hours.

## Post-publication record

- **Status:** Not checked
- **Summary:** Post-publication record check unavailable at this time.
- **Last checked:** 2026-11-25