#methodological rigourAntibody concentrations confound retrieval method comparisons
Pepsin protocols use higher antibody concentrations, including β-hCG at 1:50 versus 1:100 and vWF at 1:1,000 versus 1:3,000. The reported intensity increases therefore cannot isolate a pepsin effect, and the Discussion does not acknowledge this limitation.
↳ Table 1; Results, pepsin-intensity section; Discussion
#methodological rigourReplication and assessment safeguards remain unspecified
The donor count and inferential unit for ANOVA are not stated in the main Methods or figure legends, preventing assessment of independent replication. Blinding of the subjective cell-type grading is also unreported.
↳ Sample Collection; Image Acquisition; Morphological Analysis; Statistical Analyses; Figs. 2–4
#reportingQuantitative performance claims exceed demonstrated validation
The Abstract and Conclusion describe objective, precise or unbiased results without reporting precision estimates or validation against a reference quantification. The pepsin-intensity wording also fails to distinguish a combined protocol comparison from an isolated retrieval effect.
↳ Abstract; Results, pepsin-intensity section; Discussion; Conclusion