#methodological rigourReference comparison cannot isolate lens loss
The unlensed reference uses a separate PIC and omits an extra patch cable present in the lensed setup. Consequently, the Abstract and Conclusion's attribution of 0.85 dB per interface to lens integration is not separately established by the comparison.
↳ Abstract; Section 4.2, Fig. 3c; Section 6
#methodological rigourReplicate units and measurement precision remain undefined
The reported n = 4 and n = 3 do not explicitly distinguish independent assemblies from channels or repeated measurements, and the meaning of ± is unspecified. Alignment and extinction-ratio results also lack stated uncertainty, limiting conclusions about reproducibility.
↳ Section 4.2; Section 5.2, Figs. 3–4
#contributionPluggable connector suitability remains experimentally untested
The measured working distance and alignment tolerances establish laboratory feasibility, not validated connector operation. Mating-cycle, contamination and environmental tests are absent, while the improved lateral tolerance comes with reduced angular tolerance.
↳ Abstract; Section 4.2; Section 6