#reportingParallel demonstration exceeds the reported experiments
The abstract presents simultaneous multi-component measurement as demonstrated, but the algorithm comparison uses one connection. The Conclusions leave integration into the completed configuration to future work.
↳ Abstract; Sections 2 and 5
#methodological rigourMinimum transmission lacks verified input polarization
The Customized method infers the minimum state from two 45-degree waveplate rotations through single-mode fiber without establishing the resulting state at the device input. Paddle agreement at 1310 nm does not validate broadband PDL or the Highlights' calibrated-SOP wording, and this limitation is not discussed.
↳ Highlights; Sections 1 and 4; Figure 9
#methodological rigourUncertainty and replication remain quantitatively undefined
Independent run and device counts, repeatability estimates and a measured noise floor are absent. The 0.15 dB output spread is dismissed as noise, while stability and accuracy comparisons lack an independent calibrated reference.
↳ Sections 3–4, Figures 6 and 9