# Eulerian formulation of the tensor-based morphology equations for strain-based blood damage modeling

- **DOI:** [10.1016/j.cma.2024.116979](https://doi.org/10.1016/j.cma.2024.116979)
- **Rating:** Sound (3.7 out of 5)
- **Verdict:** An Eulerian reformulation corrects a consequential morphology-model simplification and enables practical pump calculations, but verification against an existing model does not establish biological accuracy.
- **Framework version:** v5.13
- **Pipeline version:** v0.4
- **Evaluation date:** 2026-11-26
- **Evaluation page:** https://www.nabu.science/paper/10.1016%2Fj.cma.2024.116979

## Key claims

1. Overall, this comparison demonstrates that the Eulerian formulation of the TTM is a valid approximation of the Lagrangian Arora model for the given application.
   - Support: Supported
   - Type: Primary finding
   - Basis: application-scoped agreement on selected trajectories without refinement studies establishing uniform local accuracy
   - Location: Section 4.4, end
2. [On the examined pump plane, outside the outer-wall and central-inflow regions,] the simplified model underpredicts the effective shear rate by up to 98%.
   - Support: Supported
   - Type: Secondary
   - Basis: direct comparison with TTLM on the examined plane without numerical refinement checks
   - Location: Section 4.4, Figure 8c
3. Compared to Lagrangian strain-based models, the TTM is computationally more efficient and does not require tracking individual cells.
   - Support: Tentative
   - Type: Secondary
   - Basis: absence of matched Lagrangian runtime measurements for the general efficiency comparison
   - Location: Section 5, final paragraph
4. Overall, we are confident that the TTM represents an excellent compromise between accuracy and practicality.
   - Support: Tentative
   - Type: Secondary
   - Basis: accuracy assessed against the underlying morphology model rather than experimental cell-deformation or hemolysis data
   - Location: Section 5, final paragraph

## Strengths

- **Equivalent reformulation exposes consequential predecessor errors** The derivation resolves eigenvector rotation in the Eulerian frame and explains why the earlier simplified formulation changes predicted deformation. Benchmark and pump comparisons demonstrate the practical consequences.
  - Dimension: contribution
  - Location: Sections 2.2 and 2.4, Sections 4.1–4.4
- **Verification spans canonical flows and pump applications** Direct comparisons with the originating Lagrangian model test the approximation across progressively more complex flows. In the pump, integrated results over 36 pathlines agree within 5%, although local differences remain.
  - Dimension: methodological rigour
  - Location: Sections 4.1–4.4, Figure 10
- **Biophysical limitations receive substantive literature engagement** The conclusion discusses non-ellipsoidal cell shapes and relaxation dynamics rather than treating ellipsoidal morphology as universally valid. It explicitly reserves experimental validation and more realistic device comparisons for future work.
  - Dimension: positioning
  - Location: Section 5

## Limitations

- **Numerical convergence and local accuracy remain unestablished** No mesh or time-step refinement study is reported. Four detailed pump trajectories and a 36-pathline aggregate comparison do not quantify local discretization error throughout the domain.
  - Dimension: methodological rigour
  - Location: Sections 4.3–4.4, Figure 10
- **Conflicting specifications obscure two benchmark details** The rotating-shear expression gives a midpoint angular frequency of 1 rad/s where the text states zero. The pump-plane height is reported as 0.5 mm in the text and 5 mm in Figure 8's caption.
  - Dimension: reporting
  - Location: Sections 4.2 and 4.4, Figure 8 caption
- **Practical advantage claims exceed the reported comparisons** Table 1 contains no matched Lagrangian runtime supporting the general efficiency assertion. The claimed accuracy–practicality compromise also rests on agreement with the Arora model rather than experimental cell-deformation or hemolysis validation.
  - Dimension: reporting
  - Location: Table 1, Section 5 final paragraph

## Reliability checks

Caveats (4 of 4 checks).

- **Results check:** Noted
  - "The plane is located 0.5 mm above the blades." (Section 4.4, pump-plane description) vs "5 mm above the impeller blades." (Figure 8 caption). The text and caption assign heights differing by a factor of ten to the same displayed pump plane, without identifying which is correct. The effect on the result is unclear.
  - Minor reporting notes (1):
    - "ω(r∗) = ωi + (ωo −ωi)(2r∗−1) ." (Section 4.2, angular-frequency expression) vs "ω(r∗= 0.5) = 0." (Section 4.2, rotating-shear interpretation). Substituting r∗ = 0.5 and the stated ωi = 1 rad s−1 into the printed expression gives 1 rad s−1, not zero; the implemented expression is unknown. The effect on the result is unclear.
- **Reporting and transparency:** Clean. No reporting or transparency finding warrants a conduct note. HemTracer is identified as open source, while availability of the principal Eulerian solver is not established.
  - Code availability: declared
- **Reference integrity:** Clean. 50 of 50 checkable references verified
  - 51 references in manuscript
  - 1 are books, websites or datasets — counted, but not index-checkable
  - References not verified automatically are resolved manually within 24 hours.

## Post-publication record

- **Status:** Clean
- **Summary:** No retraction or expression of concern found in Crossref (includes Retraction Watch).
- **Last checked:** 2026-11-26